Vital Signs Patient Monitor

Vendor: Contec

Description:

Features:

 8.4” color TFT display

Unique digital SPO technology with excellent measurement and interference-free

Audible and visual alarms with adjustable alarm range

Anti electrosurgical interference and defibrillation

Can be connected with LAN to work with central monitoring system

Powerful data management and storage capacity

Suitable for adult, pediatric and neonatal patient

Standard Configuration:

 ECG,HR,NIBP,SpO2,PR,RESP,2-TEMP

Optional:

  ETCO2,2-IBP, Thermal Recorder

 

Performance

▲ECG

Lead mode: 3-lead or 5-lead

Lead selection: I, II, III, aVR, aVL, aVF, V

Waveform: 5-lead, 2-channel

3-lead, 1-channel

Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV

Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s

HR: 

Measurement and alarm range: 15~350bpm

Accuracy: ±1% or ±1bpm, whichever is greater

Alarm accuracy: ± 2 bpm

Resolution: 1 bpm

ST-segment monitoring:

Measurement and alarm range: -2.0mV~+2.0mV

Accuracy: -0.8mV~+0.8mV  ±0.04mv or ±10%, whichever is greater 

Other range: unspecified 

Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

Pacemaker: yes

▲NIBP

Method: Oscillometry

Mode: Manual/AUTO/STAT 

Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes

Measurement period in STAT mode: 5 minutes

Measurement and alarm range: 10 ~ 270 mmHg

Resolution: 1 mmHg

Cuff pressure accuracy: ±3 mmHg

Measurement accuracy:

Maximal mean deviation: ±5 mmHg

Maximal standard deviation: 8mmHg

Over-pressure protection:

Adult mode: 297±3 mmHg

 Pediatric mode: 240±3 mmHg

 Neonatal mode: 147±3 mmHg

▲SpO2

Measurement and alarm range: 0 ~ 100%

Resolution: 1%

Measurement accuracy: 70%~100%: ±2%; 

0%~69%: unspecified 

PR

Measurement and alarm range: 30 ~ 250 bpm

Measurement accuracy: ±2 bpm or ±2%, whichever is greater

 

▲TEMP

Channel: dual-channel

Measurement and alarm range: 0 ~ 50℃

Resolution: 0.1 ℃

Accuracy: ±0.1 ℃

▲RESPIRATION

Method: R-F(RA-LL) Impedance 

Respiration rate: 

Measurement and alarm range: 0~150rpm

Resolution: 1 rpm

Measurement accuracy: ±2 rpm

Alarm accuracy: ±3 rpm

Apnea alarm: 10~40s

Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s