The nGenuity 8100EP1 patient monitor displays true end-tidal CO2 with numerical ETCO2 and INCO2 values as well as capnographic waveform. A smart respiration feature automatically defaults to the best available source, either CO2 or ECG TTI.
Criticare's advanced design capnometer module is accurate with breath rates up to 120 breaths per minute. Flow rate is set at 200 mL per minute for exceptionally fast response. A rise time responsiveness within 170 milliseconds provides outstanding waveform display.
The capnometer modules uses sidestream sampling and Criticare's patented WaterChek™ water traps to remove moisture without distortion to the CO2 waveform. The water traps have a large 22 mL reservoir and full occlusion clearing capability unlike many other CO2 monitoring systems.
Features and Benefits
Monitors End-tidal CO2 and inspired CO2 accurately at high respiration rates.
Gas Monitoring with full occlusion clearing and high capacity water trap.
Mini-Trends can display condensed tabular vital signs data to show recent monitoring activity.
Compatible with popular Electronic Medical Records (EMR) systems for both large facilites and small offices.
Easy-to-read 10.4" active TFT color display.
Easy to learn, easy to use -- simplified menus and dedicated function keys.
Monitor can mount quickly and easy to a wall or rolling stand
Simple-to-use, lightweight, and portable design.
Monitor comes with an integrated printer
Materials Included
8100EP1 Patient Monitor Unit
(1) SpO2 Adult Sensor
(1) Child/Small Adult BP Cuff
(1) Adult BP Cuff Quick Connect.
(1) Large Arm Adult BP Cuff.
(1) NIBP Hose - 10ft. Straight.
(1) ECG Cable Assembly.
(3) ECG Electrodes
(1) Printer
(1) Power Cord
(1) CO2 Sidestream
(1) CO2 Sensor
(5) Water Traps
(5) Sample Line
Operator's Manual
Technical Specifications
Dimensions:
Width:13"
Height:11"
Depth:10.3"
Weight:14 lbs
ECG
Number of Leads:3- or 5-lead
Lead View:User-selectable; I, II, III, aVR, aVL, aVF, V (5-lead); I, II, or III (3-lead)
Gain Selection:x1/2, x1, x2, and x4
Frequency Response: Diagnostic: 0.05 to 100 Hz (-3dB)
Monitor: 0.5 to 40 Hz (-3dB)
Electrosurgery Protection:Yes
Defibrillator Protection:Yes
Pacer Detection/Rejection:Yes
Pulse Oximetry
Range:1% to 99%
Resolution:1%
Accuracy:
70% to 99% range:±2%;
50% to 69% range:±3%;
<50%:unspecified
Method:Dual wavelength LED
Non-Invasive Blood Pressure
Technique:Oscillometric during Inflation
Inflation Pressure Range:Adult/Pediatric: 30 to 300 mmHg